How It Works Shipping Fee Terms of Use

Model RHA Humidity Sensor

Application

  • Humidity control in food processing facilities
  • Humidity control in air-conditioning equipment
  • Humidity control in pharmaceutical factories
  • Humidity control in clean rooms
  • Humidity control for musical instruments
  • Humidity control in plant factories
  • Humidity control in greenhouses/fermentation chambers
  • Humidity control in museums, sample chambers, warehouses

Features

  • A small-sized humidity sensor.
  • Outputs a linear voltage from 0.5 V to 4 V DC corresponding to the humidity level.
  • Easy to maintain connector type is used.
  • M3 screws facilitate easy mounting.

International Standards

  • Our products do not have overseas safety regulations such as CE mark, UL standard, EN standard etc.
  • Products is compliant to the RoHS directive.

* Please consult about "change of specification" and "request of specification" of the product.

Rating/Performance/Characteristics Graph

Rating/Performance

Model RHA RHA-5V
Detection Sensor Humidity
Temperature Range -10 to 60 ºC (Without dew condensation or icing)
Humidity Range 0 to 85 % RH (Without dew condensation)
Measurable Humidity Range 0 to 100 % RH (Short-time measurement without dew condensation only)
Standard Humidity Accuracy ±5 % in the RH 25 ºC (20 to 80 % RH)
Humidity Response Time 60 sec(τ 63%)
Power Supply Voltage 12 V to 24 V DC 5 V DC
Operating Voltage Range 10 V to 26.4 V DC 4.5 V to 5.5 V DC
Current Consumption 1 mA DC or less 0.5 mA or less
Resistance Output Impedance 1 kΩ
Load Resistance 100 kΩ or more
Storage Temperature Range -10 to 60 ºC (Without dew condensation or icing)
Storage Humidity Range 80 % RH or less (Without dew condensation)
Dielectric Strength 500 V AC for 1 minute (Between the live part and case)
Insulation Resistance 50 MΩ or more at 500 V DC megger (Between the live part and case)
Vibration Resistance 10 to 55 Hz, 1.5 mm double amplitude in X, Y and Z directions for 2 hours each (at power off)
Shock Resistance 200 m/s² (Approx. 20 G) in X, Y and Z directions 3 times each (at power off)
Protection IP50
Case Material Polyarylate
Connector Connector: S3B-EH(3-pin)(from J.S.T. Mfg. Co., Ltd.)
[Connections]Housing: EHR-3, Contact: SEH-001T-P0.6
(from J.S.T. Mfg. Co., Ltd.)
Weight Approx. 5 g
Options Connector harness : CNH-S3B03S26-300 (Sold Separately)

Characteristics Graph (Typical Example)

ST-RHA_EN-2.png

Outline Dimensions/Output Circuit

Outline Dimensions

ST-RHA_EN-2.png

Output Circuit

ST-RHA_EN-1.png

*1 : The 5V DC type does not have a built-in backflow prevention diode. Be sure to check the polarity of the power supply and connect it correctly.

Usage Precautions

Usage Precautions

  • When the sensor is not used, to maintain the precision, it is recommended to store it in the delivered package (in the closed anti-static bag) without opening under the following conditions :
                     Temperature range: 10 to 50 ºC
                     Humidity range: 20 to 60 % RH
    Do not use antistatic polyethylene bags (light blue, pink or rose pink color).
  • The relative humidity value greatly depends on the humidity condition.
    When measuring the surrounding temperature/humidity, be sure to confirm there is no heated part on the installation surface.
  • Use M3 pan head screws for mounting.
    The tightening torque for the case should be 0.5 N·m or less.
  • This sensor has output impedance of 1kΩ to avoid damage due to external noise.
    The impedance attenuates according to the input impedance of the signal input device.
    Use a device with as high input impedance as possible - at least 100kΩ.
  • The 5 V DC type does not have a built-in backflow prevention diode.
    Be sure to check the polarity of the power supply and connect it correctly.
  • This sensor cannot be used in dew condensation conditions.
    If this is the case, refer to the RHT series.
  • Do not use the sensor in the following environmental conditions:
    - Where there is water or moisture
    - Where there are corrosive gases
    - Where there is oil  
    - In freezing temperatures or dew condensation
    - Dusty locations
    - A place exposed to direct sunlight or radiated heat"
  • Vacuum away any serious dust contamination or else a malfunction may result.
  • For other precautions, refer to “General Precautions” for infrared temperature sensors.

*After you have put an item in your shopping cart, choose either [Quote] or [Order].
*For customized products or specifications, contact us using "Inquiry."
*Choose the currency from JPY, USD, or EUR.

Model Name Measurable Humidity Range Standard Price Inventory Type Select
RHA-5V 0 to 100 % RH
39 USD
36 EUR
STD
Model Name Measurable Humidity Range Standard Price Inventory Type Select
RHA 0 to 100 % RH
39 USD
36 EUR
BTO

STD Shipping is available immediately (on the day of order)

WIP Shipping is available within 1 to 3 weeks

BTO Shipping is available within 10 to 18 weeks

NOTE: The above information shows the typical delivery time. For actual delivery time, please confirm us as it depends on our business days and/or product quantities.

In the case of the quote request and order, please be sure to check the Terms and Conditions.
If you have made the procedure of quote request and order, it assumes that agree to the Terms and Conditions.

Quantity:
pieces pieces
  • 2 USD
    2 EUR
  • 2 USD
    2 EUR

To confirm the Request a quote / Order Cart

Go to page top